Model XD88
The X-ray Output Detector provides a means of calibrating and trouble shooting x-ray generators. The Detector output is designed to be connected directly to an oscilloscope. The x-ray waveform can be observed to find a wide variety of problems inherent in x-rays. X-ray waveforms can be documented by using a digital scope, a storage scope or an oscilloscope camera.
The detector is simply placed in the x-ray beam, then a cable is used to connect the detector to a standard oscilloscope. The resulting oscilloscope display represents the intensity-time relationship of the x-ray beam.
The detector is weighted to make it easy to place under the x-ray.
Features
- Two Year Warranty
- Optional BNC to BNC cable is available
- Weighted enclosure
- Compact Design
Applications
- Measurement of Exposure Time
- Quality Assurance
- Verification of X-ray Operation
- X-ray Calibration
- Documentation
- Troubleshooting and Repair of x-ray
- Exposure Time Calibration
- Rectifier Problems
- Contactor Problems
- Loading Characteristics
- Cable or Connector Arching
- Shutter Calibration
Specifications
OUTPUT- 200 mV peak minimum
- X-ray at 80 kVp, 100 mA, head 15 cm (6") above detector
- 100k Ohms
- Oscilloscope Impedance is typically 1 or 10 MOhm
- 51 x 32 x 20 mm (2.0 x 1.25 x .8")
- 85g (3 oz)
- via BNC to BNC cable
- XD88 - $140
- XD88cable - $36
- 2.2 meter (7.5 foot) BNC to BNC cable
- Use to connect detector to oscilloscope
Please contact ECC to order or for more information.
Sample Oscilloscope Trace
- Full Wave X-ray
- 60 kV at 50 mA
- 33 ms exposure
- Trace was taken on Link Instruments MSO-19
- Vertical - 50 mV per division
- Horizontal - 5 ms per division